How to read an AI clearance like a reference range
By Reza Motaghi

On this page
Children are under one percent of the patients in public medical imaging datasets. A 2025 census of public datasets found usable age data in 116 of them.1 In those, children were about 4,600 patients among 489,000. The children who are there skew older, with almost no infants. When the same study ran a heart-size classifier trained on adult chest films across ages, its false positives rose sharply below age eleven. Only 17 percent of cleared medical AI devices carry a paediatric label at all.
I read medical images every day. One model I tested charted a child's image with thirty-two adult teeth. A child of that age has milk teeth and adult teeth together, and thirty-two is the full count of a grown mouth. Whatever population that model was built on, the child was not in it, and nothing in its output said so. That is the whole subject of this article. A clearance is a statement about a population. The case in front of you is either in that population or it is not.
What the reference range is
Every laboratory result comes back with a range beside it, and the range is printed with the population it was measured on. Adult, child, pregnant, fasting. Nobody reads a child's value against an adult's range. A laboratory that adopts a range from a manufacturer checks it on its own patients before reporting against it. The range is not a property of the test. It is a property of the test on a named population.
An AI clearance is the same object. It says the tool met a bar on a defined, tested population. It does not say the tool was tested on the person in the chair, and the rulebook never claimed it did. The regulator's own 2025 lifecycle guidance asks a maker to describe how representative its data is of the intended population.2 It asks for the limits to be disclosed in the labelling, and for monitoring to continue after release. European law requires the training and test data of a high-risk system to be relevant and sufficiently representative for its purpose.3 The rule is not the gap. Showing the work is.
What is and is not in a set
I built the dataset my own model learns from, so I know exactly what is in it and what is not. That knowledge is the receipt. A person who has assembled a set knows three things. Every slice has a count. Some counts are too small to say anything. A case from outside the set gets an answer anyway. The default patient is whoever the set had most of, and a model does not announce when it is guessing from the neighbours.
The child with thirty-two adult teeth was not a failure of accuracy. On adults that model may have been fine. It was a failure of scope, and the scope was printed nowhere the reader could see it.
What the record shows
A 2025 census of 903 cleared AI devices in the United States found clinical performance studies reported for about half.4 A quarter stated none had been done. Of the studies that existed, one in twelve was prospective. Fewer than a third reported results by sex, and about a quarter by age. About one device in twenty was later recalled, and of the recalled devices, seven in ten had reported no clinical study at all. That last figure is an association, not a cause, and it argues that testing, where it happens, does real work.
The newest and broadest count lands harder. Across 1,357 cleared devices through the end of 2025, one in forty had a registered prospective trial.5 Fewer than one in a hundred had posted results. Three devices, in total, had ever been checked against a patient outcome such as death or readmission. A radiology-wide review found the same shape: of 717 well-documented devices, under a third had any clinical testing, while approvals accelerated.6 In one specialty the trend runs the other way.7 The share of orthopaedic devices cleared with no clinical data fell from about three in five to under one in five within a few years. A trend claim needs its specialty named, so I name it.
The subgroup picture is thinner still. A scoping review covered 545 validation studies of 252 commercial radiology products.8 It found 77 studies that reported any subgroup performance at all, covering 52 products. The share has not become more common over time. Most of the comparisons that did exist were too small to detect a gap that would matter. Two of every three tuberculosis datasets were at risk of being underpowered for a sex split. So the honest state of the evidence is not that tools perform worse on the slices they were not built for. It is that nobody counted.
Unmeasured is not the same as worse
The counter-evidence deserves its own section, because the strongest 2026 studies do not show general failure. One radiography system held its accuracy across 21,581 cases from 52 centres in twenty countries, with a quarter of the cohort children.9 It reported consistency across every demographic and geographic split. A prostate cancer model held a reasonable floor at every one of six hospitals across two scanner makers.10 It needed a site-specific threshold at each. In a mammography model checked by subgroup, breast density predicted false alarms far better than race or age.11 Race and age stopped mattering once density was controlled for.
And when deployed systems do break, the freshest longitudinal evidence says it is rarely demographics. Four clinical AI systems were tracked for up to two years after go-live.12 They degraded mainly through calibration drift and shifts in the data pipeline, not through the population changing. So the claim this article makes is epistemic, not predictive. It is not "this tool will fail on you because of who you are". It is "nobody can tell you whether it holds for you, because nobody looked". And nobody has to tell you when the tool itself changes either.
Every field certifies an envelope
An aircraft type was certified in the early 1990s against a defined range of icing conditions that excluded freezing rain and large supercooled droplets. In October 1994 one met exactly those excluded conditions over Indiana, ice built up behind the de-icing boots, and 68 people died.13 The certification was not false. The envelope had an edge, and nothing on board said it had been crossed.
A company built a hiring tool on ten years of its own hires. It found the tool had taught itself to downgrade résumés containing the word "women's". It shut the tool down rather than fix it.14 A data audit of American mortgage decisions found Black applicants denied at far higher rates than comparable white applicants.15 The gap was traced partly to a mandated credit-scoring model built on decades-old data. Its subgroup fairness had never been tested in public. In each case the system was validated on one population and met another, and the meeting was invisible from inside the system.
Three questions for any validation report
The questions are the reference range applied to a clearance. One of them I changed after reading the scoping review.
1. Which population, and how many cases per slice? Age, sex, scanner, protocol, site, indication. A slice with a count is a claim. A slice without one is a hope. A report that gives only the demographic makeup of the test set, without performance per slice, has answered a different question. Off-ramp: a tool validated on a genuinely large and diverse cohort, with per-slice results published, has answered this. Read the table instead of asking.
2. Which slices were too small to fail? My first version asked which subgroups performed worse. The review changed it: most subgroups were never measured, and most of the measured ones were underpowered. A subgroup of thirty cases cannot show a gap that would matter, so a clean result on it is not clearance, it is silence. Ask for the count behind every reassuring number. Off-ramp: where a subgroup is small because the condition is rare in it, the arithmetic changes. A rare finding needs its own count rather than a bigger sample.
3. Is the case in front of me in that set at all? A child on an adult tool. A scanner the maker never saw. An indication the study excluded. If the report cannot say, the tool has evidence about this case's neighbours, not about this case. The answer it gives is a guess with a confidence attached. Off-ramp: none. A case outside the population is outside the clearance, whatever the output says.
Then do what every laboratory does with a borrowed range. Check the tool on your own population before you report against it. Keep a local performance profile by scanner, protocol and subgroup, the way a hospital keeps a resistance profile for its own bacteria.16 The radiologists' own professional literature has asked institutions to build exactly that infrastructure for at least two years.17 The clearance opens the door. The local check decides whether the tool is used on the people who walk through it.

What the questions are not
They are not a claim that cleared tools do not work. Some have earned trust across twenty countries and both ends of the age range. The questions find out whether the one in front of you is among them.
They are not a case against clearance. In the census, recalled devices were overwhelmingly the ones that had skipped clinical testing. Testing works. The questions ask to see it.
They are not only about demographics. Scanner, protocol, site and indication are slices too, and the post-deployment record says calibration drifts before populations do. The reference range has a population and a method printed on it. Ask for both.
What I ask
I read images every day, I built a set myself, and I test models blind, against a two-radiologist reference. Before I trust a tool that arrives validated, I ask which population, how many per slice, and whether my case is in it. Most reports cannot answer the third question, and that is the finding. The answer key behind its score is itself a settled argument. The population behind that key is the range the clearance is valid for.
Read the next clearance you are shown like a reference range. If the population is not printed on it, ask for it before the case that is not in it arrives. If you build or evaluate tools that will meet cases their test set barely held, the contact page reaches me directly.
References
Footnotes
-
Lack of children in public medical imaging data points to growing age bias in biomedical AI. medRxiv preprint, June 2025, full text via PMC. 181 datasets reviewed, 116 with usable age data; 4.6 thousand children among 489 thousand patients; cardiomegaly classifier false positives rising below age 11 (r = −0.928); 17 percent of FDA-approved medical AI devices labelled for paediatric use. https://pmc.ncbi.nlm.nih.gov/articles/PMC12259205/ ↩
-
US Food and Drug Administration. Artificial intelligence-enabled device software functions: lifecycle management and marketing submission recommendations. Guidance, January 2025. https://www.fda.gov/regulatory-information/search-fda-guidance-documents/artificial-intelligence-enabled-device-software-functions-lifecycle-management-and-marketing ↩
-
Regulation (EU) 2024/1689, Article 10(3): data sets "relevant, sufficiently representative, and to the best extent possible, free of errors and complete in view of the intended purpose". https://www.euaiact.com/article/10 ↩
-
Windecker D, Baj G, Shiri I, et al. Generalizability of FDA-approved AI-enabled medical devices for clinical use. JAMA Network Open, 30 April 2025. 903 devices; studies reported for 505 (55.9 percent), none for 218 (24.1 percent); 41 prospective (8.1 percent), 12 randomised; sex-specific results in 145 studies (28.7 percent), age subgroups in 117 (23.2 percent); 43 devices (4.8 percent) recalled at a median of 1.2 years, 30 of them with no reported clinical study. https://jamanetwork.com/journals/jamanetworkopen/fullarticle/2833324 ↩
-
Abulibdeh et al. PLOS Digital Health, 19 August 2026. 1,357 FDA-cleared AI/ML devices through December 2025; 2.5 percent with a registered prospective trial, 0.9 percent with posted results, 0.2 percent (three devices) evaluated against a patient-centred outcome. https://pmc.ncbi.nlm.nih.gov/articles/PMC13489462/ ↩
-
Sivakumar R, Lue B, Kundu S. FDA approval of artificial intelligence and machine learning devices in radiology: a systematic review. JAMA Network Open, 7 November 2025. 717 devices; 29 percent with any clinical testing, 5 percent prospective; approvals from 3 percent of clearances in 1995 to 2015 to 23 percent in 2023. https://pmc.ncbi.nlm.nih.gov/articles/PMC12595527/ ↩
-
Lee B, et al. FDA-cleared artificial intelligence medical devices in orthopaedic surgery. JAAOS Global Research and Reviews, 5 February 2026. Devices with zero clinical data fell from 62.2 percent (2017 to 2019) to 17.8 percent (2022 to 2024). https://pmc.ncbi.nlm.nih.gov/articles/PMC12879955/ ↩
-
Walston SL, Takita H, Mitsuyama Y, et al. The current state of demographic subgroup reporting for commercially available AI for radiology: a scoping review. European Radiology, 2026, as reported by Medical Xpress, 2 September 2026. 545 studies, 252 products; 77 studies on 52 products with subgroup performance; 67 percent of tuberculosis datasets at risk of being underpowered for a sex subgroup. https://medicalxpress.com/news/2026-09-fragmented-bias-commercial-medical-ai.html ↩
-
Sultan JL, Beaumel P, et al. International multicenter validation of an expanded AI diagnostic system for 18 pathologies in thoracic and musculoskeletal radiography. Diagnostics, 10 April 2026. 21,581 cases, 52 centres, 20 countries, paediatric patients 24.6 percent, AUC above 96 percent. https://pmc.ncbi.nlm.nih.gov/articles/PMC13114525/ ↩
-
Giganti F, Moreira da Silva N, et al. Multi-centre, multi-scanner validation of a prostate cancer detection AI. February 2025. AUC at least 0.83 at each of six sites, 0.91 overall; site-specific thresholds recommended. https://pmc.ncbi.nlm.nih.gov/articles/PMC12226644/ ↩
-
Woo M, et al. Subgroup evaluation to understand performance gaps in deep learning-based classification of regions of interest on mammography. PLOS Digital Health, 8 April 2025. Dense-breast categories 1.9 to 2.5 times the false-positive risk; race and age not significant once density was controlled for. https://pmc.ncbi.nlm.nih.gov/articles/PMC11978028/ ↩
-
Kopanitsa G. Validation is not enough: longitudinal evidence of post-deployment fragility in clinical AI systems. PLOS Digital Health, 27 July 2026. Four systems, 18 to 24 months; calibration drift and workflow or data-pipeline shifts dominated. https://journals.plos.org/digitalhealth/article?id=10.1371/journal.pdig.0001534 ↩
-
American Eagle Flight 4184, 31 October 1994, as summarised from the NTSB's report AAR-96/01: the icing certification envelope excluded freezing rain and supercooled large droplets; ice accreted behind the de-icing boots; 68 killed. Tertiary summary; the NTSB report itself was not opened for this article. https://en.wikipedia.org/wiki/American_Eagle_Flight_4184 ↩
-
Algorithmic bias, summarising Reuters reporting of October 2018 on a résumé-screening tool built on ten years of one company's hires. Tertiary summary. https://en.wikipedia.org/wiki/Algorithmic_bias ↩
-
The Markup. The secret bias hidden in mortgage-approval algorithms. 25 August 2021. Black applicants denied at 80 percent higher rates and Latino applicants at 40 percent higher rates than comparable white applicants in 2019 federal data. https://themarkup.org/denied/2021/08/25/the-secret-bias-hidden-in-mortgage-approval-algorithms ↩
-
From plug-and-play to institution-calibrated radiology AI. Frontiers in Radiology, 16 July 2026. An institution-level performance profile modelled on the microbiology antibiogram, stratified by scanner, protocol, indication and subgroup. https://www.frontiersin.org/journals/radiology/articles/10.3389/fradi.2026.1902217/full ↩
-
Ramwala O, et al. Establishing a validation infrastructure for imaging-based artificial intelligence algorithms before clinical implementation. Journal of the American College of Radiology, 22 May 2024. https://pmc.ncbi.nlm.nih.gov/articles/PMC11486600/ ↩